ANESTHESIA TROLLEY / BOYLES APPARATUS

Trolley Type Anaesthesia Trolley
Structure Material Mild Steel
Work Top Material Mild Steel

Feature

Dual pressure gauges for accurate monitoring of oxygen and nitrous oxide supply
Oxygen failure alarm ensures patient safety with audible warning during low pressure
Supports multiple vaporizers for flexible anesthesia delivery
Emergency oxygen flush provides instant high-flow oxygen when required
Built-in safety system prevents nitrous oxide delivery without oxygen pressure

Technical Specification

Pressure Gauges: Two each for O₂ & N₂O cylinders pressure and one each for O₂ & N₂O line pressure
Oxygen Failure Warning Device: Pneumatic device which activates audible alarm (for minimum 10 seconds) when Oxygen supply pressure falls to 205 kPa.
Vaporizers: Space for three Vaporizers provided as per standard.
Oxygen Flush: Emergency Oxygen flush outlet button at table level. (36–50 lit./min.)
Change Over Unit: Change over unit provided for open or close circuit.
Nitrous Cut-off: Nitrous cut-off system prevents risk of administering 100% Nitrous Oxide in absence of Oxygen pressure.
All Accessories are optional
Packing:

Height: Approx. 1400 mm
Width: Approx. 600 mm
Depth: Approx. 500 mm
Weight: ~60 kg