Pump Type INFUSION PUMP
Channel Type Single Channel
Accuracy ±5%
Alarm Type Occlusion
Display Type LCD
Power Source Rechargeable Battery
Performance
Rate: 1 ml/h ~ 1,200 ml/h
Flow Rate Accuracy: Within ±5% (after correct calibration)
Mechanical Precision: Within ±2%
Purge Rate: 100 ml/h ~ 1,000 ml/h (100 ml/h step)
Infusion Volume: 1 ml ~ 9,999 ml
Total Infusion Volume: 0.1 ml ~ 9,999.9 ml
KVO Rate: 1 ml/h ~ 5 ml/h (1 ml/h step)
Occlusion: High: 800 mmHg ±200 mmHg (106.7 kPa ±26.7 kPa) Medium: 500 mmHg ±100 mmHg (66.7 kPa ±13.3 kPa) Low: 300 mmHg ±100 mmHg (40.7 kPa ±13.3 kPa)
Unique Audible and Visible Alarm: Human voice alarm for end of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out, etc.
Power Source:
AC 100V ~ 240V, 50/60Hz
Internal rechargeable Li battery, capacity ≥1600 mAh
Internal battery backup
Bubble Detector:
Ultrasonic wave detector
Detection sensitivity ≥25 μL
Fuse: F1AL / 250V, 2 pcs inside
Power Consumption: 18 VA
Operating Condition:
Ambient temperature: +5°C ~ +40°C
Relative humidity: 20 ~ 90%
Atmospheric pressure: 86.0 kPa ~ 106.0 kPa
Transport & Storage Condition:
Ambient temperature: −30°C ~ +55°C
Relative humidity: ≤95%
Equipment Classification: Class II, internal power supply, Type CF
IP Classification: IPX4
Performance
Rate: 1 ml/h ~ 1,200 ml/h
Flow Rate Accuracy: Within ±5% (after correct calibration)
Mechanical Precision: Within ±2%
Purge Rate: 100 ml/h ~ 1,000 ml/h (100 ml/h step)
Infusion Volume: 1 ml ~ 9,999 ml
Total Infusion Volume: 0.1 ml ~ 9,999.9 ml
KVO Rate: 1 ml/h ~ 5 ml/h (1 ml/h step)
Occlusion: High: 800 mmHg ±200 mmHg (106.7 kPa ±26.7 kPa) Medium: 500 mmHg ±100 mmHg (66.7 kPa ±13.3 kPa) Low: 300 mmHg ±100 mmHg (40.7 kPa ±13.3 kPa)
Unique Audible and Visible Alarm: Human voice alarm for end of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out, etc.
Power Source:
AC 100V ~ 240V, 50/60Hz
Internal rechargeable Li battery, capacity ≥1600 mAh
Internal battery backup
Bubble Detector:
Ultrasonic wave detector
Detection sensitivity ≥25 μL
Fuse: F1AL / 250V, 2 pcs inside
Power Consumption: 18 VA
Operating Condition:
Ambient temperature: +5°C ~ +40°C
Relative humidity: 20 ~ 90%
Atmospheric pressure: 86.0 kPa ~ 106.0 kPa
Transport & Storage Condition:
Ambient temperature: −30°C ~ +55°C
Relative humidity: ≤95%
Equipment Classification: Class II, internal power supply, Type CF
IP Classification: IPX4
Feature
HD LCD Display, high capacity words, friendly user interface, dynamically display working status;
Audible and visual alarm for occlusion, empty, low battery, end of infusion, door open, wrong setting, etc., which gain patents;
Compatible with any brands of infusion sets after correct calibration;
Preset solution volume to greatly reduce the workload of nurses;
Work mode: ml/h and drop/min can switch freely;
Three levels of occlusion: high, middle, and low;
Purge function;
KVO (keep-vein-open) automatically opens as the infusion is completed, KVO rate is 1-5ml/h(1ml/h step);
Power Source: AC100—240V, 50/60Hz; Internal Battery;
Automatically record the settings of the last infusion;