SYRINGE PUMP

Channel Type Single Channel
Flow Rate Range 0.1 to 1500 ml/hr
Syringe Size 5 ml
Accuracy +-1 %
Occlusion Level High
Display Type LCD
Power Source Rechargeable Battery

Feature

HD LCD Display, high capacity words, friendly user interface, dynamically display working status;
Audible and visual alarm for occlusion, empty, near empty, low battery, end of infusion, syringe loose, wrong setting, etc.;
Compatible with 5ml, 10ml, 20ml, 30ml, 50/60ml syringe of any brands;
Preset solution volume to greatly reduce the workload of nurses;
Three working modes: Rate mode, Time Volume mode, Dosage Weight mode;
Three levels of occlusion: high, medium and low;
Purge and Bolus function;
KVO (keep-vein-open) automatically opens as injection is completed, KVO rate is
1-5ml/h(0.1ml/h step);
Freely Stackable: freely stack one syringe pump onto another to provide multiple
solutions, which have a wide range of clinical applications;
Power Source: AC100—240V, 50/60Hz; Internal Battery, DC12V car charge;
One-key operation makes setup easy and simple;
Syringe plunger grabble detector, operable with one hand in a No-Germ Environment;
Automatically record the settings of last injection

Packaging

Dimension: 280mm (L), 210mm(W),130mm(H)
Weight: 2.2kg (net weight)
Performance
Unique Human Voice Alarm system
The device triggered the human voice prompt and stop
injection automatically when there is any malfunction,
making the infusion process safer and more reliable.
Rate:
50/60ml Syringe: 0.1ml/h ~ 999.9ml/h (0.1ml/h step)
1,000ml/h ~ 1,500ml/h (1ml/h step)
30ml Syringe: 0.1ml/h ~ 900.0ml/h (0.1ml/h step)
20ml Syringe: 0.1ml/h ~ 600.0ml/h (0.1ml/h step)
10ml Syringe: 0.1ml/h ~ 300.0ml/h (0.1ml/h step)
Bolus Rate:
50/60ml Syringe: 1,200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h

Volume Limit – 0.1ml ~ 9999.9ml (0.1ml step)
Total Injection Volume – 0.1ml ~ 9999.9ml (0.1ml step)
Occlusion of Single Channel Syringe Pump
High: 800mmHg +/-200mmHg (106.7kPa +/-26.7kPa)
Medium: 50/600mmHg +/-100mmHg (66.7kPa +/-13.3kPa)
Low: 300mmHg +/-100mmHg (40.7kPa +/-13.3kPa )

KVO rate: 1ml/h
Alarm:
Injection soon finish, end of injection, occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.